Biochemistry Analyzer | Labmedica Expo
Resumen
|
Productos
CARETIUM
Biochemistry Analyzer  
Modelo: KB-120
Consulta Rápida
Descripción
The KB-120 fully automatic biochemistry analyzer has a throughput of up to 50- 120 tests per hour and features separate sample probe & reagent probe to guarantee precision of sample volume/reagent volume. The sample probe & reagent probe are equipped with liquid level detection feature and automatic cleaning (inside & outside), while the system offers automatic recording of all operation steps and reaction curves.

Información Requerida
Solicite una Quota

Caracteristicas
1. Separate sample probe & reagent probe in order to guarantee the precision of sample volume / reagent volume.
2. Sample probe & reagent probe are equipped with liquid level detection feature and automatic cleaning (inside & outside)
3. Automatic recording of all operation steps and reaction curves
4. High quality distribution device with 5 years shelf life without any maintenance required, the sampling aspiration is as accurate as 0.1µl for the sample probe.
5. Real-time display of residual reagent volume, operational in mono and bi-reagents.
6. Displaying the working state of the sample tray and reaction cuvettes
7. Ability to run STAT samples at any moment.
8. Automatic re-testing of high-concentration results
9. Automatic cleaning of reaction cuvettes
10. Throughput : 50- 120 tests/h
11. Lab information management system has the review / editing / checking / analyse and print functions for the test results

Especificación
• Optics system
Light source : halogen lamp 12v/20w
Interference filter : 340nm、405nm、492nm、510nm、546nm、578nm 、630nm
Bandwidth : 8 nm~12nm
Absorbance range : -0.3000~2.5000ABS
Absorbance precision : 0.0001ABS
Stray light : <1.0%(340nm)
• Sample system
Sample position : 40 positions (including positions for calibrations, controls and cleaning solutions)
Sample volume range: 2.0~45.0µl . Precision : 0.1µl
Sample probe : automatic liquid surface detection
Sample probe cleaning : full cleaning of internal and outer probe surfaces
• Reagent system
Reagent volume range : 20.0~700.0µl, precision: 0.5µl
Reagent positions : 20 positions
Storage capacity of reagent bottles : 25~40ml
Reagent probe : automatic liquid surface detection
Reagent probe cleaning : full cleaning of internal and outer probe surfaces
• Reaction and analysis
No. of Reaction cuvettes : 112, Auto-washable
Final volume required for reading : 300-500µl
Flow cell : 1.0cm ,32µl , Duo flow cells
Temperature of flow cell : 37℃, precision ±0.1℃
Carry over : <2.0%
Analysis methods : end-point , two-point , kinetic
Analysis speed : 50~120 tests/hour
Calibration method : one point or multi-point linear calibration , multi-point non-linear calibration
Calibration type : calibration after start-up
Quality control : 2 groups of quality control data with different concentration, and displaying L-J quality control graph and related statistic data.
• Operation system
Windows 2000 operation system with Chinese and English interfaces
Display : LCD screen
Printing function: Setting to print full testing report, quality control graphs and working volume statistic reports etc.
Interface : Standard RS-232
• Working conditions
Power supply : 220V±10%,50Hz
Temperature : 10℃~40℃
Humidity : 20%~85%
Water consumption : 5L/H
Dimension: 880 mm × 580mm × 580mm (L×W×H)
Weight: 60 kg


Información Requerida
Solicite una Quota